How To Calibrate a Digital Scale

Every digital scale has a different sequence of buttons you need to press to calibrate them

If you find or suspect that your scale is not working properly and you need to calibrate it we have scale quite a few scale calibration procedures listed here.  Most scales will need a know calibration weight.  Typically you can calibrate a scale at both zero (dead load).  And Span which is the known weight portion of the procedure.  So before you proceed with any of these instructions please have a known weight ready.  These instructions are for reference only.  Progressive Scale takes no responsibility for the use of these instructions.

If you need an Onsite Scale Calibration in the CT, NY, MA area please call 203-792-2854.  Scale Systems.   Scale Software.  Meat Processing Software.  Label Printing Scales

Home
About
Contact
How To Calibrate a Digital Scale

A&D A & D AD 4321 Calibration Procedure for A & D AD 4321 Balance 
Turn off Turn throw switch in back press zero CAL1 press stby oper and you should get ................. in display enter the calibration calibration weight using the keys place calibration weight on scale and press stby oper you should get.................. slide switch back take off calibration weight You be done
weight scales



A&D A&D EK 6000H Calibration Procedure for A&D EK 6000H Balance 
HOLD SAMPLE AND PRINT CAL OUT CAL 0 PRESS SAMPLE SHOWS Cal press print cal 0 prees print add calibration calibration weights press print end remove calibration weight done
weight scales



A&D A&D EP 60 KB Calibration Procedure for A&D EP 60 KB Balance 
A&D EP 60 KB Throw switch on side cal 0 press print Enter Cal Wt press tare (possibly twice) calF put on calibration weight press tare Cal End
weight scales



A&D AND FV-30K Calibration Procedure for AND FV-30K Balance 
OPEN BACK OF SCALE PUSH SWITCH TO "ON" PRESS "UNITS" DISPLAY WILL READ "CAL O" PRESS "ZERO" PRESS UNITS DISPLAY READS "CAL 3" WAIT FOR CIRCLE IN UPPER LEFT CORNOR PRESS "ZERO" DISPLAY WILL READ "END" SLIDE SWITCH BACK
weight scales



A&D AND HV or FW-60K Calibration Procedure for AND HV or FW-60K Balance 
(FV-30K 60LB/40LB)(FW-10K) CAL 1: MAX CAP. IN KG = 150/60/30KG. CAL 2: 2/3 CAP. IN KG = 100/40/20KG. CAL 3: MAX CAP. IN LB = 300/120/60LBS. CAL 4: 2/3 CAP. IN LB = 200/80/40LBS. OPEN UP BATTERY COMPARTMENT REMOVE SCREW FLIP CAL SWITCH TO ON IT WILL READ 9.798 PRESS "MODE/UNITS" "CAL O" STABLE LIGHT PRESS "REZERO" "CAL1" PRESS "MODE' TO SELECT SETTING CAL 1,2,3,4 (HIGH/LOW/GO WILL ROTATE SETTINGS) PUT calibration weight ON (200LBS.) AND WAIT FOR STABLE LIGHT PRESS "REZERO" "CAL END" IS DISPLAYED PUT SWITCH BACK
weight scales



Acculab ACCULAB S50 AND S100 Calibration Procedure for ACCULAB S50 AND S100 Balance 
PRESS MODE TO "KG" MODE PRESS AND HOLD "CAL" UNTIL CAL. WT. MODE APPEARS PRESS "TARE" (WHILE CAL IS STILL FLASHING) "CAL O" WILL BE DISPLAYED PAUSE 2 SECONDS PLACE CORRECT CAL. calibration weight ON SCALE "CAL F" DONE WHEN DISPLAY READS CORRECT calibration weight PRESS "TARE" TO RESET ZERO
weight scales



Acculab Acculab VA8000 Calibration Procedure for Acculab VA8000 Balance 
Hold F1 & on /off self test -c- press tare f1 cal displayed put on calibration weight 5000g turn off
weight scales



Acculab Acculab Balance Calibration Procedure for Acculab Balance Balance 
Empty Scale Press and hold Cal/mode until calibration calibration weight appears in display Put calibration weight on scale Wait until a beep. Tthe cal calibration weight will blink and it will go back into weigh mode
weight scales



Adam Equipment TFW 60 Calibration Procedure for TFW 60 Bench Scale 
Turn the Power OffTurn the power back on , during the counting from 9 to 0 press func/c key. The display will show "CAL" for a few seconds. While it is showing "CAL" press the SMPL,PRINT and TARE keys in sequence to enter the calibration section. The display will show "unload". Remove any calibration weight from the platform. Press ZERO/ENTER key. The display will show the last calibration calibration weight used. If this is correct you cn continue by pressing the ZERO/ENTER key. If it is not correct use the arrow keys to change the calibration calibration weight value. When it is correct press the ZERO/ENTER key. If the calibration is acceptable the displaywill show "PASS" and then retur to normal. If an error message is shown try calibration again as a disturbance may have prevented a successful calibration.
weight scales



Cardinal Cardinal 2235 Calibration Procedure for Cardinal 2235 Counting Scale 
Remove right side plate& throw switch Press Scale Select for scale being calibrated Change Calib)? to 1 Press count button till Load___ Load Scale Enter calibration calibration weight Press Count Unload Scale Press Count Till Done
weight scales



Denver Denver 3000D Calibration Procedure for Denver 3000D Balance 
hold tare till cal put on 500 or 1000 or 2000 or 3000 in gram calibration weights cal x cal done
weight scales



Detecto Detecto 2225 Calibration Procedure for Detecto 2225 Counting Scale 
Turn Scale Off Remove right cover plate Remove blue Jumper in the corner of the scale attached to the display board (don't confuse with jumper on main board) Power on scale 7670? press1 then count cap= (whatever) choose remote or local now if remote just press count now. If local press scale select then count Press count button till Load___ Load Scale Enter calibration weight Press Count Unload Scale Press Count Till Done 
weight scales



Digi Digi DI-20 & 28 Calibration Procedure for Digi DI-20 & 28 calibration weight Scale 
For the 20 the Jumper JP-2 must be enabled on the main board Press And hold the rezero button and press the up button to the right of the tare button then tare twice It then displays the scale increments change with arrow keys or accept and push rezero Enter the capacity of the scale push rezero Enter the calibration weight you will use to cal push rezero cal 0 appears push rezero with nothing on scale cal sp appears add calibration weight and press rezero then press tare to get out
weight scales



Digi Digi DC-120, DC-130  , DC-190 IQ9500 Calibration Procedure for Digi DC-120,130 ,190 Counting Scale 
select scale to calibrate using" scale" button Press and hold the rezero key and press 8715 let go of rezero empty scale press rezero put calibration weight on scale enter amount of cal wt press "*" press mode twice do this per scale attached
weight scales



Digi Digi DC-150 Calibration Procedure for Digi DC-150 Counting Scale 
Press mode repeatedly until maintenance. Press and hold rezero then press 8715. release rezero press tare (right arrow key) till span adjust press enter the left of the display is the scale number to change this press + to the right of that is the scale capacity. To the riqht of that is the calibration calibration weight. The calibration calibration weight has to be entered with the same amount of decimals as the scale capacity. To change either number use the tare to highiite then enter new ualue. Once all values here are ok press enter. Empty scale push rezero add calibration weight push enter * press mode till operation press enter 
weight scales



Digi Diqi DC-60 Calibration Procedure for Diqi DC-60 Counting Scale 
Remove label casette In the back of opening there is a 3 position switch. Move this switch into the middle position. Select scale to calibrate using scale button. Turn key to program Press and hold rezero while pressing 8715 release rezero. Empty scale push rezero add calibration weight enter calibration weight push enter turn key back to operate move switch back to original position 
weight scales



Doran Doran 8000 Calibration Procedure for Doran 8000 Bench Scale 
Doran 8000 Remove back cover plate Throw red switch inside Displays AUGAI Press "conv" Display ENT 0 Put calibration weight on scale zero=up tare = down to correct calibration weight Press "conv" displays ent sp Take calibration weight off flip switch back
weight scales



Doran Doran 8000XL Calibration Procedure for Doran 8000XL Bench Scale 
Doran 8000 xl press cal sw if cap ok press units cal 0 press zero put calibration weight on scale press units press print hold print to go lower or capicity to go higher till cal calibration weight is matched press zero done



Doran 4200 Calibration Procedure for 4200 Bench Scale 
remove small plate in back of scale. thow red switch press gross net c000.00 press zero press units add calibration weight use arrows to correct units to save gross net 
weight scales



DORAN 7000 DORAN Calibration Procedure for DORAN Bench Scale 
OPEN UP SQUARE IN BACK (RECTANGULAR) "FLIP RED SWITCH" "CO" PRESS "REZERO" (COUNTS DOWN) C=calibration weight VALUE PUT THAT AMOUNT ON PRESS "REZERO" (COUNTS DOWN) C=calibration weight VALUE PUT THAT calibration weight ON PRESS "REZERO" (COUNTS DOWN) DONE FLIP SWITCH BACK 
weight scales



Fairbanks Fairbanks HR-2300-2 Calibration Procedure for Fairbanks HR-2300-2 calibration weight Scale 
Fairbanks HR-2300-2 Press Program press ^ till cal code is 113 then enter wait for scale I forgot from there 
weight scales



Fairbanks Fairbanks UPS Calibration Procedure for Fairbanks UPS calibration weight Scale 
Fairbanks ups scale remove plate on left side press button press zero press units place displayed calibration weight on scale press cal button
weight scales



Fairbanks Eclipse PPI-HR5000-X Calibration Procedure for Eclipse PPI-HR5000-X Bench Scale 
remove back cover press black button on board twice to get to CC CAL Hit units when reads 0 add calibration weight. Hit units again. Adjust to correct calibration weight using zero/AT sw on side of unit then hit units again
weight scales



Filing Filing DR-2100A Calibration Procedure for Filing DR-2100A Fork Truck Scale 
Filing DR2100A 9090 ENTER DEL TOGGLES Y N MENU ENTER FOLLOW INST IF CAL SW OFF OPEN BOX AND THROW SWITCH
weight scales
Filing Filing DR-5000 Calibration Procedure for Filing DR-5000 Fork Truck Scale 
Filing DR-5000 The F1 & F2 buttons are shut off points for filling applications Cal Procedure To get into calibration mode press "9","9","Re-Zero" This calls the true zero. Go to the pot labeled R-40 and get as close as possibleto zero Push zero button Apply load Go to R-41 and dial to correct calibration weight Remove calibration weight Press "Re-zero" Re adjust if needed Once that is accomplished Call up true zero again. Go to R-40 and turn to negative 1/2 the scale capacity (-2500 to -2600lb) then rezero with button re apply calibration weight if incorrect, turn fine span pot 1/4 clockwise for up or 1/4 counter clock wise for down Do not watch the indcator because this will screw you up. Re zero the scale and re apply the calibration weight. repeat the procedure until it is correct. 



Flexweigh DWM IV Calibration Procedure for DWM IV Indicator 
Throw SW1 ->FSE F1 Push 5 FSE F6 Press Print F6.0 0 Print Print Print f6.3d Tareprint F6.40 print Escale print capacity shows, press enter add load after countdown add load push print enter calibration weight push enter
weight scales



GSC Scale GSC Scale Calibration Procedure for GSC Scale Indicator 
Turn On, press ¨Stanby¨, after press¨ZERO¨hold and press ¨STANBY¨Display will ready ¨SYS¨. Press ¨ZERO¨, display will ready 00000, the las digit will be blinked, Press ¨P¨ to increase to 6, press ¨NET/GROSS¨ to move to the left, press ¨P¨to increase to 5, press ¨NET/GROSS¨ to move to the left, press ¨P¨ to increase to 4, the display will read 00456, Press ¨ZERO¨, the display will show F1, press ¨SET¨,the display will show ¨F0¨, press ¨SET¨, the display will show CAL, press ¨ZERO¨, remove load from the balance, press ¨NET¨, Press ¨P¨ to include the value of weigths that will be use for calibration. Put Load on the scale and press¨ZERO¨, wait just a moment. DONE.
weight scales



GSE GSE-350 Calibration Procedure for GSE-350 calibration weight Scale 
GSE 350 Indicator Press zero and select simultaneously press zero,print,units and tare one after the other scale should say first zero. Press tare with nothing on the scale use th print and units keys to enter the load calibration weight press tare Add calibration weight press tare press tare till out 
weight scales



GSE 450 GSE 460 Calibration Procedure for GSE 450 460 calibration weight Scale 
GSE 450 & 460 Indicator 1) Press [Zero]+[Select] Simultaneously 2) Press [Zero] [PRINT] [UNITS] [ENTER] In that order 3) Remove any extraneous Load 4) Press[Select] until "New Zero" is dislayed 5) Press[Enter] 6) place cal calibration weight on scale 7) Key in the calibration weight value 8) Press [Enter] "Cal OK" 9) Press [Enter] 3 times 
weight scales



GSE 455 GSE  465 GSE  550 Calibration Procedure for GSE 455 465 550 calibration weight Scale 
GSE 455-465-550-574-650-660 1) From the calibration weight screen keyy in 100 [Select] DIsplays Key in Code: The access code must be keyed in within 5 seconds 2) Press 54321 [ID] [Enter] 3) Press Enter 4) Apply cal calibration weight 5) Key in calibration weight and press enter 6) Cal OK Press [Enter] 7) Enter=Save press [Enter] 8) Enter=Exit [Enter]
weight scales



GSE  625 622 Calibration Procedure for GSE 625 622 calibration weight Scale 
GSE 625 or 622 Counting Scale Use the scale select key to choose the scale you wish to calibrate Press the [GSE][MODE] to put the instrument into offset display mode press the [GSE][ZERO]. The scale is now in the auto cal mode Before going any further enter the program jumper bypass code [23640][GSE][.] The lower display will prompt you through the set up of Auto Cal. First of all remove all calibration weight and press [Zero]. The scale will average the zero while it counts to 10 in the lower display. Counter will reseet when motion is sensed When tho count reaches 10 place the calibration weight on scale,enter that amount and press[ENTER] When the procedure is complete press[MODE] to return to offset displayed If you want to cal another scale use scale select and start all over or press [ZERO] to re-enter the Weigh Mode
weight scales



Masstron  M -1500 Calibration Procedure for Masstron M -1500 calibration weight Scale 
Masstron M1500 Open Indicator and there is a fine span pot labeled R-17 Little black pot in middle of board Medium span SW3 M5000 Open indicator and main group of boards is on a hinge. undo screw in back so boards swing out. Then throw SW!-1 on the back board. then on keypad type "access" "0" "5"then press enter this is the zero cal. astrange number is there press enter till done. For span press "access" "0" "4" then enter. place calibration weights on scale. tell the indicator what it is then press enter. To save changes press access then enter twice then wait two seconds before throwing SW1-1 back. 
weight scales



NCI Counting Scale Calibration Procedure for NCI Counting Scale Counting Scale 
NCI Counting scale Go to scale to calibrate Take off cover remove small cover plate throw sw2 on s2 to closed put on 1 calibration weight enter value and press acc+ Do 2nd value the same way then throw back switch
weight scales



Ohaus  DS-10 Calibration Procedure for Ohaus DS-10 Balance 
Ohaus DS-10 110 LB Capacity Hold button for 15-20 sec 0 Cal in window place 110 LB on scale it will say 50 KG Remove calibration weight
weight scales
Pennsylvania Pennsylvania 2720 Calibration Procedure for Pennsylvania 2720 Counting Scale 
PENNSYLVANIA 2720 20 LB CAPACITY SCALE: PRESS "2" PRESS "ENTER" ADD .5 THEN PUT ON 10 LBS. PRESS "ENTER" ADD 1 THEN PUT ON 20 LBS. PRESS "ENTER" ............GOES BLANK THEN TAKE calibration weight OFF PRESS (KEY IN) "20" ALL ZEROES INCLUDED PRESS "ENTER" PRESS ".005" PRESS "ENTER" (*SAME BELOW FOR SCALE 2) PRESS "2.20462" PRESS "ENTER" TURN KEY BACK SCALE 2 100 LB. CAPACITY: PRESS "2" PRESS "ENTER" ADD .5 PUT ON 50 LBS THEN "ENTER" ADD 1 PUT ON 100LBS. THEN "ENTER" .......... TAKE calibration weight OFF PRESS "100.00" THEN "ENTER" GO BACK TO * 
weight scales



Pennsylvania Pennsylvania 420-10 Calibration Procedure for Pennsylvania 420-10 Counting Scale 
PENNSYLVANIA 420-10 (10/200 CAP) SAMPLE=SMALL PLATFORM BULK=LARGE PLATFORM TURN KEY TO PROGRAM KEY IN SENSITIVITY OF LOAD CELL "2" THEN "ENTER" ADD .5 (1/2 CAP. OR KEY IN OTHER calibration weight) PUT calibration weight ON PRESS "ENTER" ADD 1 (FULL CAPACITY OR ENTER IN AMOUNT) PUT calibration weight ON PRESS "ENTER" .......WILL STROBE AND THEN BLANK OUT TAKE calibration weight OFF .......WILL STROBE AGAIN AND THEN BLANK OUT (POT SCREW TO 1.5 MV OR LIFT UP CORNOR OF PLATFORM) KEY IN CAPACITY OF LOAD CELL INCLUDING ALL ZEROS (2OO.OO) PRESS "ENTER" BLANK OUT .05 RESOLUTION KEY IN PRESS "ENTER" ONLY FOR SAMPLE SCALE: PRESS "2.20462" PRESS "ENTER" 
weight scales



Pennsylvania Pennsylvania 5400 Calibration Procedure for Pennsylvania 5400 Counting Scale 
Pennsylvania 5400 Counting 1)Turn Key in Back of scale 2) Enter scale number to be calibrated 3) Press enter till cal 18 STOP 4) Wait about 10 sec 5) Press Enter 6) Press enter again 7) Empty scale and press enter 8) Put 1/2 scale capacity on scale and press enter OR put different calibration weight on scale and enter value then press enter 9) Put Full scale capacity calibration weight on scale or enter value and press enter 10) Press eter till STOP 11) Turn Key Bac k
weight scales



Pennsylvania Pennsylvania 5600 Calibration Procedure for Pennsylvania 5600 calibration weight Scale 
Penn 5600 hold down 4 key full capicity press enter till cal type use units to change to load cell press enter follow instructions or enter different calibration weight values followed by enter
weight scales



Pennsylvania Pennsylvania 7500 Calibration Procedure for Pennsylvania 7500 Bench Scale 
Penn 7500 push button under plate under platter. The plate is held in place by a screw on the bottom of the scale. Push the button under the cover. Press print till cal 42 press print to cal zero. Put on half capacity and press print. Cal 44 put on full capacity and press print. Cal zero print . then print to cal 99 and the scale will count down and all done.
weight scales



Rice Lake Rice Lake CW-40 Calibration Procedure for Rice Lake CW-40 Bench Scale 
Rice Lake CW-40 Flip Red Switch behind access plate Press under button Display C press zero for zero cal scale says ent zero add calibration weight and adjust with up and down arrows. Press Units Flip switch back
weight scales



Rice Lake Rice Lake CW-80
Calibration procedure for Rice Lake CW-80 Bench Scale

Press config Switch.  Right over to calibration.  Scroll down to zero. Clear the platter and press the zero .  will display zcal for 10 sec. when done calsp. display shifts to full scale cap weight. adjust weight and press zero.  scal is displayed while working. CalIB means it is done. remove weights. Press config switch. you are done

Rice Lake Rice Lake IQ-600HB Calibration Procedure for Rice Lake IQ-600HB Weigh Scale 
Rice Lake IQ 600 HB TAKE DOWN METAL BAR IN FRONT 4 SWITCHES FLIP SWITCH 3 1234 (GET READING AS CLOSE TO ZERO W/PENTIAMETER) FLIP SWITCH BACK PRESS "ZERO" CLOSE SWITCH #1 (MAY NOT BE ZERO THAT`S OK) PRESS "ZERO" WITH NO calibration weight ON "CAL O" "C" PLACE 10% ON SCALE PRESS (KEY IN) 500 LBS. (NET/GROSS =UP TARE=DOWN) PRESS "CONV" FLIP SWITCH BACK 



Rice Lake IQ 6500 6200 9500 Calibration Procedure for IQ 6500 6200 9500 Counting Scale 
Rice ake IQ 6200, 6500, 9500 Press and hold the rezero key and press 8715 let go of rezero press rezero put calibration weight on scale enter amount of cal wt press "*" press mode twice do this per scale attached



Rice Lake IQ 355 Calibration Procedure for IQ 355 Weighing Scale 
Remove the large Philips or Flat Head Screw on the back of the indicator.  Using a straightened paper clip poke in the hole where the screw was.  Indicator will read "ConFiG" Press Right arrow (PRINT) twice to "CALibr"  Remove all weight from the scale. Press Down arrow (Gross/Net) to "WZEro" Press Down (Gross/NET) The display show "0CAL0" for a moment then it will display some numbers.  Press TARE to "WVAL".  Press down (Gross/Net) .  You will enter the calibration weight here.  To select the digit you want to change us the left and right arrows (UNITS or PRINT).  Use the up and down arrows (GROSS/NET or ZERO) to change the digit value. Once the correct calibration weight is being displayed press the TARE button to "WSPAn"  Put Known Weight on the scale.  Press the down arrow (GROSS/NET).  Scale will display "0CAL0" for a moment then a large number.  Press TARE.  Press the up arrow (ZERO) until "CALibr", Press Left arrow (UNITS) till "ConFiG" then press up arrow (ZERO) You are calibrated.





​Rice Lake IQ 710 Calibration Procedure for IQ 710 Weighing Scale 
Remove the large Philips or Flat Head Screw on the back of the indicator.  Using a straightened paper clip poke in the hole where the screw was.  Indicator will read "ConFiG" Press Right arrow (PRINT) twice to "CALibr"  Remove all weight from the scale. Press Down arrow (Gross/Net) to "WZEro" Press Down (Gross/NET) The display show "0CAL0" for a moment then it will display some numbers.  Press TARE to "WVAL".  Press down (Gross/Net) .  You will enter the calibration weight here.  To select the digit you want to change us the left and right arrows (UNITS or PRINT).  Use the up and down arrows (GROSS/NET or ZERO) to change the digit value. Once the correct calibration weight is being displayed press the TARE button to "WSPAn"  Put Known Weight on the scale.  Press the down arrow (GROSS/NET).  Scale will display "0CAL0" for a moment then a large number.  Press TARE.  Press the up arrow (ZERO) until "CALibr", Press Left arrow (UNITS) till "ConFiG" then press up arrow (ZERO) You are calibrated.





Rice Lake 920i Calibration Procedure for Rice Lake 920i Weight Scale System
Remove the large Philips screw on the bottom left of the scale indicator (Older Indicators Flat Head) .  This will reveal a 1/4 inch hole.  Use a ball point pen and press in poke into the hole and this will put the scale into configuration mode.  [SCALES] will be selected.  Using the arrows press DOWN to SCALE1.  If this is the scale you wish to calibrate press DOWN otherwise select the scale using the arrows and press DOWN. [GRADS] should be selected.  LEFT arrow to [CALIBR].  DOWN to [WZERO].  Remove  all weight from the scale and press the DOWN arrow.  Press the "Calibrate" button.  Indicator will say Calibrating Dead Load... Wait for Deadload Calibration to come back then press UP arrow to [WZERO] then RIGHT arrow to [WVAL].  Press DOWN arrow.  Using the numeric keypad, enter the calibration weight you are going to place on the scale then press "enter". [WSPAN] should be selected.  Place known weight on scale and press the DOWN arrow.  Press the "Calibrate" button.  The scale will say Calibrating Span...Once it comes back to Span Calibration you are done.  Press the UP arrow back till [SCALES] then press "Save and Exit"


Rice Lake Rice Lake RS-130 Calibration Procedure for Rice Lake RS-130 click here
Sartorius CP3000 Calibration Procedure for CP3000 Balance 
Hold Down Tare Apply calibration weight Displayed Follow Instructions
weight scales





Rinstrum R320, X320, CAS X320 Calibration. Press SELECT first and hold then press UNITS/Power for 2 seconds at the same time.  The screen will go to [BUILD]
Press ZERO key 2 times to get [CAL]
Press the TARE key 1 time to get [ZERO]
Remove all weight from the platform 
Press the SELECT key one time to get weight flashing
Press the PRINT key one time [2 in 1 P] then [0.00]
Press TARE key one time to get [ZERO]
Press TARE key one time to get [SPAN]
Press SELECT key one time (if you get CLR then press the f (OK) key)
Add calibration weight to the scale
To edit the weight
SELECT will select the digit · PRINT will edit the digit (scroll up)
Press the f (OK) key to get 5in P then the weight will show
To save Press the select and power again and hold till saving.

Rinstrum R420 Calibration. Unplug scale and power back up. Press and hold [power] and [f3]. Scale will say GEN.OPT. Press [ZERO]key twice to get SCALE. Press [TARE] 3 times to get CAL. Press [SELECT]. Get ZERO CAL ?. Press [OK] on right keypad . Remove weight from the scale. Press [OK] get Z in P (Zero in Progress). Wait till you get 0.000 DONE. Press [OK] get ZERO CAL ?. Press [SELECT] key to get SPAN CAL ?. Press [OK]. Put Weight on Scale. Press [OK]. Enter the weight on the scale. Press [OK]. Get S IN P
Get weight and DONE. Press [OK]. Get SPAN CAL ?. Press [Power]. You are Done 

Scientech Scientech Calibration Procedure for Scientech Bench Scale 
Scientech Balance 1) Press the FUNCTION switch (or MDE and RANGE switches simultaneously for dual range models) 2) Press the MODE switch when CAL1 appears. 3) Wait 10 seconds. Then press the ZERO switch 4) Place one of the flashing calibration calibration weights on the pan 5) ait 10 seconds. Then press the zero switch 6) Remove calibration weight 
weight scales



Setra Calibration Procedure for Setra Counting Scale 
Setra counting scales Press "-" press"clear" press "zero" put calibration weight on scale enter calibration weight value press "set up/cal"
weight scales



Sterling Sterling Calibration Procedure for Sterling Counting Scale 
Sterling 825 Counting Scale 1)Press Scale Select to go to the scale you are calibrating 2) Press and hold Gross/Net Key until Params 3) Now the NTEP parameter along with its value is displayed 4)Press the print key or zero key till FCAL 5)Select FCALT with tare key and press Print 6) Shows current number f scales to change tis pram press tare or press print to move on 7) This is capacity change it or press print to move on 8)Graduations press print or change 9)Shows Cal Units print to move on 10) CAL F Change this to CAL T using Tare button press print to begin cal a)Clear scale and press print b)Step A8 press print c)Place calibration weight on scale, enter calibration weight value and press print 11)SPN 0.0 this is the span error press print 12)Shows caacity of next scale. Repeat steps 7-11 13)Gross/Net or reset key will get you out at the end
weight scales



Toledo Toledo 8571 Calibration Procedure for Toledo 8571 Counting Scale 
TOLEDO 8571 PRESS "ENTER" UNTIL E SCALE (F 3.2 F 3.3) AFTER E SCALE PRESS "*" (PRINT) ADD LD (TARE=CHANGE #'S CLEAR=NEXT # ZERO=BACKS UP) "OOO" ADD calibration weight PRESS "*" (PRINT) PRESS "CLEAR" PRESS "*PRINT" UNTIL F 4.5 THEN "*" (PRINT) SET OFF 
weight scales



Toledo Toledo 8581 Calibration Procedure for Toledo 8581 calibration weight Scale 
Toledo 8581 Counting Scale Throw Switch under platter Press 0 to F3 Press print to F3.2 Press1 Scale Cap Displayed Press Print Increment Size Press Print E Scl (Empty Scale) Press Print Scale Counts Down Add calibration weight Press Print Enter calibration weight in whole numbers Press Print Counts Down Cal d (Done) Press print till set up off Throw switch back Turn scale off & on
weight scales



Toledo Toledo SC-05 Calibration Procedure for Toledo SC-05 calibration weight Scale 
Toledo SC-05 lift platter. Press cal switch in hole to the right of platter. Puts yuu into service. Use rezero enter and tare to get into cal mode
weight scales



Transcell TI-500E Calibration Procedure for TI-500E Scale Indicator 
Turn Off Scale. Throw Switch on back of scale. Turn On Scale. F1 press right arrow till F16. Press Down Arrow. C0 Press Zero Key. Press Net/Gross. Says Set. F17 with right arrows. Arrow down. C1 then flash 0. Add calibration weight. Change weigh using arrow buttons. Press Gross/Net end C1 Turn off and throw switch back 
weight scales



Vibra Screw Filling System Calibration Procedure for Filling System Filling System 
Throw cal switch on dip switches located at the bottom of aluminum box. Press mode to get to gross mode press the 8 button (also says cal). Displays SPCd press enter enter 5321 then enter setup press enter till cal press enter Hd Cal Press enter "Zero" displays press enter "-" displays press test clr 0.0 displays press enter span displays place calibration weight on scale press enter "-" displays press test/clr enter calibration weight press enter lincor press enter 0.0 displays press enter return displays press enter end cal displays press enter switch back